MPX Series Magnetostrictive Level Sensors Rev. B, 4/13
32
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]
Out of Range Samples (1 to 255) (40414)
Determines the number of consecuve level readings that fall outside of the
acceptance window before the “out of range” target is recognized as being
legimate and gured into the averaging buer. Out of Range readings
detected fewer mes than the Out of Range Samples seng will be rejected.
Sample Rate (50 to 1000 msec) (40415)
Sets the interval between oat readings. Opons allow rates from 50 to
1000 msec. A higher Sample Rate will result in a more rapid response to level
changes. Lowering the Sample Rate will help decrease the chance of detecng
errant readings. It is highly recommended that the Sample Rate be set only as
fast as is necessary for the applicaon. A sample rate of 500 to 1000 msec. is
usually appropriate for slow moving level sensing applicaons.
Mulplier (0 to 19999) (40416)
Sets the calibraon mulplier that will be applied to the sensor readings.
Oset (-10364 to 10364 mm) (40417)
Used to adjust the zero reference point of the sensor.
RTD Oset (-100 to 100 mm) (40421)
Used to calibrate the internal temperature sensor located at the boom of the
stem (MPX-E1 and MPX-R1 only).
Float Window (0 to 1000 mm; 0 = single oat) (40422)
Sets the distance from top oat to the point where the sensor will start looking
for a second oat. If the Float Window is set too small, the trailing end of
the top oat signal will be picked up as the boom oat level, causing errant
readings.
NOTE: the Float Window should always be set to 0 when using only 1 oat.
1st and 2nd Float Oset (-10364 to 10364 mm) (40423 to 40424)
Used to calibrate the top and boom oat readings.
Gain Oset (0 to 255) (40425)
Factory set--do not adjust. Default = 128.