MPX Series Magnetostrictive Level Sensors Rev. B, 4/13
32
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Out of Range Samples (1 to 255) (40414)
Determines the number of consecuve level readings that fall outside of the
acceptance window before the “out of range” target is recognized as being
legimate and gured into the averaging buer. Out of Range readings
detected fewer mes than the Out of Range Samples seng will be rejected.
Sample Rate (50 to 1000 msec) (40415)
Sets the interval between oat readings. Opons allow rates from 50 to
1000 msec. A higher Sample Rate will result in a more rapid response to level
changes. Lowering the Sample Rate will help decrease the chance of detecng
errant readings. It is highly recommended that the Sample Rate be set only as
fast as is necessary for the applicaon. A sample rate of 500 to 1000 msec. is
usually appropriate for slow moving level sensing applicaons.
Mulplier (0 to 19999) (40416)
Sets the calibraon mulplier that will be applied to the sensor readings.
Oset (-10364 to 10364 mm) (40417)
Used to adjust the zero reference point of the sensor.
RTD Oset (-100 to 100 mm) (40421)
Used to calibrate the internal temperature sensor located at the boom of the
stem (MPX-E1 and MPX-R1 only).
Float Window (0 to 1000 mm; 0 = single oat) (40422)
Sets the distance from top oat to the point where the sensor will start looking
for a second oat. If the Float Window is set too small, the trailing end of
the top oat signal will be picked up as the boom oat level, causing errant
readings.
NOTE: the Float Window should always be set to 0 when using only 1 oat.
1st and 2nd Float Oset (-10364 to 10364 mm) (40423 to 40424)
Used to calibrate the top and boom oat readings.
Gain Oset (0 to 255) (40425)
Factory set--do not adjust. Default = 128.